Test Code C8FX C8 Complement, Functional, Serum
Reporting Name
C8 Complement, Functional, SUseful For
Diagnosis of C8 deficiency
Investigation of a patient with an undetectable total hemolytic complement level
Method Name
Automated Liposome Lysis Assay
Performing Laboratory
Mayo Clinic Laboratories in RochesterSpecimen Type
Serum RedOrdering Guidance
The total complement assay (COM / Complement, Total, Serum) should be used as a screen for suspected complement deficiencies before ordering individual complement component assays. A deficiency of an individual component of the complement cascade will result in an undetectable total complement level.
Specimen Required
Patient Preparation: Fasting preferred
Supplies: Sarstedt 5 mL Aliquot Tube (T914)
Collection Container/Tube: Red top
Submission Container/Tube: Plastic vial
Specimen Volume: 1 mL
Collection Instructions:
1. Immediately after specimen collection, place the tube on wet ice.
2. Centrifuge and aliquot serum into plastic vial.
3. Immediately freeze specimen.
Specimen Minimum Volume
0.5 mL
Specimen Stability Information
Specimen Type | Temperature | Time | Special Container |
---|---|---|---|
Serum Red | Frozen | 14 days |
Reject Due To
Gross hemolysis | OK |
Gross lipemia | Reject |
Gross icterus | OK |
Reference Values
33-58 U/mL
Day(s) Performed
Monday through FridayCPT Code Information
86161
LOINC Code Information
Test ID | Test Order Name | Order LOINC Value |
---|---|---|
C8FX | C8 Complement, Functional, S | 50997-6 |
Result ID | Test Result Name | Result LOINC Value |
---|---|---|
C8FX | C8 Complement, Functional, S | 50997-6 |
Clinical Information
Complement proteins are components of the innate immune system. There are 3 pathways to complement activation: 1) the classical pathway, 2) the alternative (or properdin) pathway, and 3) the lectin (mannan-binding lectin) pathway. The classical pathway of the complement system is composed of a series of proteins that are activated in response to the presence of immune complexes. A single IgM molecule or 2 IgG molecules are sufficient to trigger activation of the recognition complex initiated by C1q. The activation process triggers a cascade that includes an amplification loop. The amplification loop is mediated by C3, with cleavage of a series of proteins, and results in 3 main end products: 1) anaphylatoxins that promote inflammation (C3a, C5a), 2) opsonization peptides that are chemotactic for neutrophils (C3b) and facilitate phagocytosis, and 3) the membrane attack complex (MAC), which promotes cell lysis.
Patients with deficiencies of the late complement proteins (C5, C6, C7, C8, and C9) are unable to form the MAC, and may have increased susceptibility to neisserial infections.
C8 deficiency is relatively rare, over 50 cases have been described. The C8 protein is comprised of 3 subunits: alpha, beta, and gamma. However, variants leading to deficiency have not been reported in C8 gamma, and the majority are in the C8 beta subunit. C8 deficiency is characterized by recurrent neisserial infections, particularly meningitis. Autoimmune disease (systemic lupus erythematosus-like) has also been reported. Given the 3 subunits, it is possible to have a low-normal C8 concentration but a nonfunctional protein, therefore the recommendation for testing is the functional assay.
For most of the complement proteins, a small number of cases have been described in which the protein is present but is nonfunctional. These rare cases require a functional assay to detect the deficiency.
Interpretation
Low levels of complement may be due to inherited deficiencies, acquired deficiencies, or due to complement consumption (eg, as a consequence of infectious or autoimmune processes).
Absent C8 levels in the presence of normal C3 and C4 values are consistent with a C8 deficiency. Absent C8 levels in the presence of low C3 and C4 values suggests complement consumption.
Normal results indicate both normal C8 protein levels and normal functional activity.
Cautions
Absent (or low) C8 functional levels in the presence of normal C8 antigen levels should be replicated with a new serum specimen to confirm that C8 inactivation did not occur during shipping.
Method Description
C8 complement activity is measured by mixing patient serum with a C8-deficient serum. The lytic activity of the serum mixture is tested against sensitized, labeled liposomes. If lysis occurs, the patient serum must be the source of the C8. The target liposomes are a commercial reagent (WAKO total complement CH50), and the assay is performed on an Advia XPT.(Unpublished Mayo method)