Test Code CATLN Calcitonin, Fine-Needle Aspiration Biopsy Needle Wash, Lymph Node
Reporting Name
Calcitonin, FNAB, Lymph NodeUseful For
As an adjunct to cytologic examination of fine-needle aspiration specimens in athyrotic individuals treated for medullary thyroid carcinoma to confirm or exclude metastases in enlarged or ultrasonographically suspicious lymph nodes
Method Name
Electrochemiluminescence Immunoassay
Performing Laboratory
Mayo Clinic Laboratories in RochesterSpecimen Type
Fine Needle WashShipping Instructions
Send specimen frozen.
Necessary Information
The biopsied site of each specimen is required and must be clearly identified in the LIS and/or batch sheet.
Specimen Required
Patient Preparation: For 12 hours before this procedure do not take multivitamins or dietary supplements containing biotin (vitamin B7), which is commonly found in hair, skin, and nail supplements and multivitamins.
Container/Tube: Plain, plastic, screw-top tube
Specimen Volume: 1 to 1.5 mL
Collection Instructions:
1. Needle wash specimens for analysis should be collected in conjunction with cytology specimens.
2. Have saline available prior to start of procedure. Saline is the only acceptable solution for needle washings.
3. After each fine-needle aspiration biopsy (FNAB) has been collected and the material in the needle has been expelled onto a slide for cytologic analysis, attach the used FNAB needle to an empty syringe.
4. Withdraw between 0.10 mL and 0.25 mL of saline up through the needle until the saline starts to fill the hub of the needle or end of the syringe.
5. Expel this fluid back through the needle into a separate tube. This is the needle washing used for analysis.
6. Repeat steps 2 through 4 for each needle pass of the same biopsied site and empty into the same tube, accumulating a total of 0.5 mL to 1.5 mL of fluid to send to the laboratory. (If more than 1 site is biopsied, see Additional Information)
7. Inspect specimen for visible blood or tissue contamination:
a. If bloody, centrifuge specimen and transfer supernatant to a new plastic aliquot tube (5-mL standard tube) to send to laboratory. The supernatant, not the cellular material, is used for analysis.
b. If specimen is clear, centrifugation is not necessary.
8. Freeze within 2 to 4 hours of collection.
Additional Information:
1. If more than 1 site is biopsied, each washing material should be submitted in a separate tube and under a different order number.
2. A minimum of 0.5 mL is required for testing; however, the total collection volume should not exceed 1.5 mL. Sample volumes outside these parameters may be rejected.
3. Do not send saline control. This test has been validated to rule-out saline matrix effect.
Specimen Minimum Volume
0.5 mL
Specimen Stability Information
Specimen Type | Temperature | Time | Special Container |
---|---|---|---|
Fine Needle Wash | Frozen (preferred) | 7 days | |
Refrigerated | 4 hours |
Reject Due To
Gross hemolysis | Reject |
Gross icterus | OK |
Reference Values
An interpretive report will be provided.
Day(s) Performed
Monday through Saturday
CPT Code Information
82308
LOINC Code Information
Test ID | Test Order Name | Order LOINC Value |
---|---|---|
CATLN | Calcitonin, FNAB, Lymph Node | 75709-6 |
Result ID | Test Result Name | Result LOINC Value |
---|---|---|
CATL | Calcitonin, FNAB, Lymph Node | 75709-6 |
SITEB | Site | 39111-0 |
Clinical Information
Calcitonin is a polypeptide hormone secreted by the parafollicular cells (also referred to as calcitonin cells or C-cells) of the thyroid gland. Malignant tumors arising from thyroid C-cells (medullary thyroid carcinoma: MTC) usually produce elevated levels of calcitonin. MTC is an uncommon malignant thyroid tumor, comprising less than 5% of all thyroid malignancies. Measurement of serum calcitonin is used in the follow-up of patients who underwent surgical removal of the thyroid gland.
Studies have reported that the measurement of calcitonin in fine-needle aspiration biopsy (FNAB)-needle washes improves the evaluation of suspicious lymph nodes in patients with a history of MTC when used in combination with cytology. Comparing the results of calcitonin in the needle rinse with serum calcitonin is highly recommended. An elevated calcitonin in the serum could falsely elevate calcitonin in the washings if the rinse is contaminated with blood. In these cases, only calcitonin values significantly higher than the serum should be considered as true-positive results.
Cytologic examination and measurement of calcitonin can be performed on the same specimen. To measure calcitonin, the FNA needle is rinsed with a small volume of normal saline solution immediately after a specimen for cytological examination (for a smear or CytoTrap preparation) has been expelled from the needle. Calcitonin levels are measured in the needle wash.
Interpretation
In athyrotic patients with a history of medullary thyroid carcinoma, a fine-needle aspiration calcitonin value greater than or equal to 5.0 pg/mL is suggestive of the presence of metastatic MTC in the biopsied lymph node.
Calcitonin values less than 5.0 pg/mL suggest the lymph node does not contain medullary thyroid carcinoma. This result is dependent on accurate sampling and a total needle wash volume between 0.5 to 1.5 mL.
This test should be interpreted in the context of the clinical presentation, imaging, and cytology findings. If the results are discordant with the clinical presentation, a sampling error at the time of biopsy should be considered.
Cautions
Samples should not be taken from patients receiving therapy with high biotin or vitamin B7 doses (ie, >5 mg/day) until at least 12 hours following the last biotin administration.
Blood contamination during the biopsy might lead to false elevations of calcitonin in the fine-needle aspiration biopsy needle washing if serum calcitonin is significantly elevated. If blood was present in the washing, only calcitonin values significantly higher than the serum should be considered as true positive results.
In some immunoassays, the presence of unusually high concentrations of analyte may result in a high-dose "hook" effect. This may result in a lower or even normal measured analyte concentration. If the reported result is inconsistent with the clinical presentation, the laboratory should be alerted for troubleshooting.
In rare cases, some individuals can develop antibodies to mouse or other animal antibodies (often referred to as human anti-mouse antibodies [HAMA] or heterophile antibodies), which may cause interference in some immunoassays. The presence of antibodies to streptavidin or ruthenium can also rarely occur and may also interfere in this assay. Caution should be used in interpretation of results and the laboratory should be alerted if the result does not correlate with the clinical presentation.
Results are dependent on accurate sampling and a maximum needle wash volume of 1.5 mL or less.
While the needle washes from several distinct needle passes or aspirations from a single area should be pooled, biopsies from different areas should be submitted as separate specimens.
Method Description
The Roche human calcitonin (hCT) assay is a sandwich, electrochemiluminescence immunoassay that employs a biotinylated monoclonal hCT-specific antibody and a monoclonal hCT-specific antibody. Calcitonin in the specimen reacts with both the biotinylated monoclonal hCT-specific antibody and the monoclonal hCT-specific antibody labeled with a ruthenium complex, forming a sandwich complex. Streptavidin-coated microparticles are added, and the mixture aspirated into the measuring cell where the microparticles are magnetically captured onto the surface of the electrode. Unbound substances are then removed with ProCell. Application of voltage to the electrode induces the chemiluminescent emission, which is then measured.(Package insert: Elecsys Calcitonin. Roche Diagnostics; 12/2021)
For all samples with high concentrations of hCT, a dilution series is performed. A linear dilution excludes hooking and most major interferences. Samples that contain low hCT concentrations are spiked with exogenous hCT to identify possible interferences that may cause a false-low result.