Sign in →

Test Code CDU Cadmium, 24 Hour, Urine

Reporting Name

Cadmium, 24 Hr, U

Useful For

Detecting exposure to cadmium, a toxic heavy metal, in 24-hour urine specimens

Method Name

Triple-Quadrupole Inductively Coupled Plasma Mass Spectrometry (ICP-MS/MS)

Performing Laboratory

Mayo Clinic Laboratories in Rochester

Specimen Type

Urine


Ordering Guidance


If employees are being monitored in the workplace, OSHA requires that laboratory reports express the cadmium excretion rate per gram of creatinine rather than per 24 hours. Order CDUOE / Cadmium, Occupational Exposure, Random, Urine to accommodate that requirement. Mayo Clinic Laboratories is certified to provide this test.



Necessary Information


24-Hour volume (in milliliters) is required.



Specimen Required


Patient Preparation:

1. For the 48-hour period prior to start of collection, as well as during the collection, patient should not eat seafood.

2. High concentrations of gadolinium and iodine are known to interfere with most metal tests. If either gadolinium- or iodine-containing contrast media has been administered, a specimen should not be collected for 96 hours.

Supplies: Urine Tubes, 10 mL (T068)

Collection Container/Tube: Clean, plastic urine container with no metal cap or glued insert

Submission Container/Tube: Plastic, 10-mL urine tube or clean, plastic aliquot container with no metal cap or glued insert

Specimen Volume: 3 mL

Collection Instructions:

1. Collect urine for 24 hours.

2. Refrigerate specimen within 4 hours of completion of 24-hour collection.

3. See Metals Analysis Specimen Collection and Transport for complete instructions.

Additional Information: See Urine Preservatives-Collection and Transportation for 24-Hour Urine Specimens for multiple collections.


Specimen Minimum Volume

1.5 mL

Specimen Stability Information

Specimen Type Temperature Time Special Container
Urine Refrigerated (preferred) 28 days
  Ambient  28 days
  Frozen  28 days

Reject Due To

  All specimens will be evaluated at Mayo Clinic Laboratories for test suitability.

Reference Values

0-17 years: Not established

≥18 years: <0.7 mcg/24 h

Day(s) Performed

Monday through Friday

CPT Code Information

82300

LOINC Code Information

Test ID Test Order Name Order LOINC Value
CDU Cadmium, 24 Hr, U 5612-7

 

Result ID Test Result Name Result LOINC Value
31106 Cadmium, 24 Hr, U 5612-7
TIME7 Collection Duration 13362-9
VL83 Total Volume 3167-4

Clinical Information

The toxicity of cadmium resembles the other heavy metals (arsenic, mercury, and lead) in that it attacks the kidney; kidney dysfunction with proteinuria with slow onset (over a period of years) is the typical presentation. Measurable changes in proximal tubule function, such as decreased clearance of para-aminohippuric acid also occur over a period of years and precede overt kidney failure.

 

Breathing the fumes of cadmium vapors leads to nasal epithelial deterioration and pulmonary congestion resembling chronic emphysema.

 

For nonsmokers, the primary source of cadmium exposure is from the food supply. In general, leafy vegetables such as lettuce and spinach, potatoes and grains, peanuts, soybeans, and sunflower seeds contain high levels of cadmium. For smokers, the most common source of cadmium exposure is tobacco smoke, which has been implicated as the primary sources of the metal leading to reproductive toxicity in both males and females.

 

The concentration of cadmium in the kidneys and in the urine is elevated in some patients exposed to cadmium.

Interpretation

Urine cadmium levels primarily reflect total body burden of cadmium. Cadmium excretion above 3.0 mcg/g creatinine indicates significant exposure to cadmium.

 

For occupational testing, OSHA cadmium standard is less than 3.0 mcg/g creatinine and, the biological exposure index is 5 mcg/g creatinine.

 

Collection of urine over 24 hours minimizes fluctuations of observed cadmium concentrations in random urine samples.

Cautions

Collection of urine specimens through a catheter frequently results in elevated values because rubber contains trace amounts of cadmium that are extracted as urine passes through the catheter.

Method Description

The metal of interest is analyzed by triple-quadrupole inductively coupled plasma mass spectrometry.(Unpublished Mayo method)

Report Available

1 to 3 days

Test Classification

This test was developed and its performance characteristics determined by Mayo Clinic in a manner consistent with CLIA requirements. It has not been cleared or approved by the US Food and Drug Administration.