Sign in →

Test Code ECMP Eculizumab Monitoring Panel, Serum


Ordering Guidance


To measure only serum concentration of eculizumab, order ECULI / Eculizumab, Serum.



Specimen Required


Patient Preparation:

1. Fasting preferred.

2. Suggest discontinuing natalizumab at least 4 weeks prior to testing for eculizumab quantitation in serum. Patient should consult the healthcare provider who prescribed this drug to determine if discontinuation is an option. If not, ok to proceed with testing while taking natalizumab.

Supplies: Sarstedt Aliquot Tube, 5 mL (T914)

Collection Container/Tube: Red top (serum gel/SST are not acceptable)

Submission Container/Tube: 2 Plastic vials

Specimen Volume: 2 mL in 2 plastic vials, each vial containing 1 mL

Collection Instructions:

1. Draw blood immediately before next scheduled dose.

2. Immediately after specimen collection, place the tube on wet ice.

3. After sample has clotted on wet ice, centrifuge at 4° C and aliquot serum into two 5 mL plastic vials.

4. Freeze specimen within 30 minutes of centrifugation. Sample must be placed on dry ice if not frozen immediately.


Useful For

Monitoring of complement blockage by eculizumab

 

Assessing the response to eculizumab therapy

 

Assessing the need for dose escalation

 

Evaluating the potential for dose de-escalation or discontinuation of therapy in remission states

 

Monitoring patients who need to be above a certain eculizumab concentration to improve the odds of a clinical response for therapy optimization

 

This test is not useful as the sole basis for a diagnosis or treatment decisions.

Profile Information

Test ID Reporting Name Available Separately Always Performed
ECULI Eculizumab, S Yes Yes
EAH50 Eculizumab Complement Blockage, S No Yes
ECUIN Eculizumab Interpretation, S No Yes

Method Name

EAH50: Enzyme-Linked Immunosorbent Assay (ELISA)

ECULI: Liquid Chromatography Tandem Mass Spectrometry (LC-MS/MS), High Resolution Accurate Mass

Reporting Name

Eculizumab Monitoring Panel, S

Specimen Type

Serum
Serum Red

Specimen Minimum Volume

1 mL total in 2 plastic vials, each vial containing 0.5 mL

Specimen Stability Information

Specimen Type Temperature Time Special Container
Serum Frozen 14 days
Serum Red Frozen 14 days

Reject Due To

Gross hemolysis Reject
Gross lipemia Reject
Gross icterus OK

Clinical Information

Eculizumab (Soliris, Alexion Pharmaceuticals), a humanized monoclonal IgG2/4 kappa antibody therapeutic directed against complement component C5, has been heralded as a breakthrough treatment for paroxysmal nocturnal hemoglobinuria (PNH) and atypical hemolytic uremic syndrome (aHUS). More recently, eculizumab has been approved to treat aquaporin-4 (AQP4) IgG positive neuromyelitis optica spectrum disorder and generalized myasthenia gravis. By association with C5, eculizumab inhibits the terminal complement pathway through simultaneous blockade of the generation of the potent prothrombotic and proinflammatory molecule, C5a, and the formation of membrane attack complex initiator, C5b.

 

Eculizumab is administered as an intravenous infusion and the dosing regimen prescribed for an average adult diagnosed with PNH is 600 mg weekly for the first 4 weeks, followed by 900 mg for the fifth dose 1 week later, then 900 mg every 2 weeks thereafter. Eculizumab has been evaluated in aHUS patients through 2 prospective, open-label, single-arm studies (C08-002 and C08-003) as well as a single-arm retrospective study. In aHUS, it is prescribed for an average adult at 900 mg weekly for the first 4 weeks, followed by 1200 mg for the fifth dose 1 week later, then 1200 mg every 2 weeks thereafter. Eculizumab was generally well tolerated, and no significant adverse effects were attributed to drug treatment; some adverse reactions included upper respiratory tract infections and diarrhea in prospective and retrospective studies, hypertension, headache, and leucopenia (C08-002/C08-003), and fever (C09-001R). Additional case reports suggest that eculizumab may prevent post-transplantation recurrence of aHUS, even in those patients harboring CFH/CFHR1 hybrid gene variants who are at very high risk of recurrence. Further research is needed to determine the duration of eculizumab therapy in the context of the genetic background of aHUS cases and risk of disease relapse.

 

Therapeutic drug monitoring of eculizumab is helpful when providers are considering personalized treatment decisions such as therapy discontinuation or extending dose intervals when patients are in remission states. In PNH, a minimum therapeutic concentration is expected to be above 35 mcg/mL and in aHUS, the therapeutic concentrations are expected to be above 50 to 100 mcg/mL of eculizumab. Complement blockage studies can aid in determining that a therapeutic concentration of the drug has blocked the complement function and subsequent production of sC5b-9. Here we offer a panel of eculizumab concentration plus alternative pathway function to monitor eculizumab therapy efficacy.

Reference Values

Eculizumab Complement Blockage:

≥46% normal

 

Eculizumab:

Lower limit of quantitation =5.0 mcg/mL

>35 Therapeutic concentration for paroxysmal nocturnal hemoglobinuria (PNH)

>50 Therapeutic concentration for atypical hemolytic uremic syndrome (aHUS)

Interpretation

Minimum trough therapeutic concentrations (immediately before next infusion) of eculizumab are expected to be above 35 mcg/mL for paroxysmal nocturnal hemoglobinuria and above 50 mcg/mL for atypical hemolytic uremic syndrome.

 

For the complement blockage monitoring of eculizumab:

-When eculizumab is present in serum at concentrations around 100 mcg/mL, the results are below the limit of quantitation of the assay (<10% of normal).

Cautions

Eculizumab complement blockage monitoring is a functional test and is dependent on correct sampling, storage, and shipping conditions. Both degradation by temperature and consumption of complement components will lead to false low function results. These are difficult to differentiate from real complement dysregulation or blockage, and in the event of poor pre-analytical handling, eculizumab concentrations are a more reliable indicator, as not subject to stringent temperature stability.

 

While pre-analytic handling can lead to falsely low results, it is far less likely that it would lead to falsely normal results.

 

Complement testing may be ordered in several circumstances where standard treatment includes plasmapheresis or plasma exchange. The procedure itself, if traumatic, may activate complement and therefore, may not be a true reflection of the patient's complement system. The recommendation is to collect blood prior to the plasma exchange whenever possible.

 

Functional results inconsistent with the clinical history should be verified with a new blood draw.

 

Specimens should be frozen immediately after collection.

 

Long term stability is optimal when the sample is kept at -70° Celsius or lower prior to testing.

 

Results must be interpreted within the clinical context of the patient.

 

Patients in transition between eculizumab and ravulizumab administration will have a result that is the sum of eculizumab plus ravulizumab in circulation. This assay will not clearly differentiate between these specific analytes and must be interpreted with caution.

 

Patients actively undergoing therapy with both natalizumab and eculizumab (extremely rare scenario) could present as assay interference. It is suggested patients discuss with their doctors the possibility of discontinuing natalizumab 4 weeks prior to testing. If discontinuation is not possible, it is ok to proceed with testing.

Method Description

The Wieslab enzyme-linked immunosorbent assay (ELISA) complement assay for the alternative pathway combines principles of the hemolytic assay for complement activation with the use of labeled antibodies specific for neoantigens produced as a result of complement activation. The microtiter plate strips are coated with lipopolysaccharide. Patient serum is diluted in diluent containing specific blocker to ensure that only the alternative pathway is activated. During the first incubation, the diluted patient serum in the wells is activated by the coating. The wells are then washed and C5b-9 (membrane attack complex: MAC) is detected with a specific alkaline phosphatase labeled antibody to the neoantigen expressed during MAC formation. After a final wash, an alkaline phosphatase substrate is added. The amount of alternative pathway complement activity correlates with the color intensity of the solution and is measured in terms of absorbance (optical density).(Nordin JG, Truedsson L, Sjoholm A. New procedure for detection of complement deficiency by ELISA. Analysis of activation pathways and circumvention of rheumatoid factor influence. J Immunol Methods. 1993;166(2):263-270; Frazer-Abel A, Sepiashvili L, Mbughuni MM, Willrich MA. Overview of laboratory testing and clinical presentations of complement deficiencies and dysregulation. Adv Clin Chem. 2016;77:1-75. doi:10.1016/bs.acc.2016.06.001)

 

Eculizumab is extracted from serum and measured by liquid chromatography (high-resolution accurate-mass) mass spectrometry.(Unpublished Mayo method)

Day(s) Performed

Varies

Report Available

3 to 12 days

Performing Laboratory

Mayo Clinic Laboratories in Rochester

Test Classification

This test was developed and its performance characteristics determined by Mayo Clinic in a manner consistent with CLIA requirements. It has not been cleared or approved by the US Food and Drug Administration.

CPT Code Information

80299

86161

LOINC Code Information

Test ID Test Order Name Order LOINC Value
ECMP Eculizumab Monitoring Panel, S 101922-3

 

Result ID Test Result Name Result LOINC Value
65676 Eculizumab, S 90240-3
619951 Eculizumab Interpretation 59462-2
618697 Eculizumab Complement Blockage, S 74520-8