Test Code GLICP CD8 T-Cell Immune Competence Panel, Global, Whole Blood
Reporting Name
CD8 Immune Competence PanelUseful For
Determining over immunosuppression within the CD8 T-cell compartment, when used on transplant recipients and patients with autoimmune disorders receiving therapy with immunosuppressant agents
Profile Information
Test ID | Reporting Name | Available Separately | Always Performed |
---|---|---|---|
TBBS | QN Lymphocyte Subsets: T, B, and NK | Yes | Yes |
GLIC | CD8 Immune Competence, B | Yes | Yes |
Testing Algorithm
When multiple specimen types are required to perform a panel of tests, the laboratory will perform the tests for which the appropriate specimen type was received, and the laboratory will cancel those for which the appropriate specimen was not received.
Method Name
Flow Cytometry
Performing Laboratory
Mayo Clinic Laboratories in RochesterSpecimen Type
WB Sodium HeparinWhole Blood EDTA
Ordering Guidance
This test requires multiple whole blood specimens to perform all testing. If only one whole blood specimen type is received, only the testing associated with that specimen type will be performed. Be advised that this may change the degree of interpretation received with the report.
Shipping Instructions
Testing is performed Monday through Friday. Specimens not received by 4 p.m. (CST) on Friday may be canceled.
Collect and package specimens as close to shipping time as possible. Ship specimens overnight in an Ambient Shipping Box-Critical Specimens Only (T668) following the instructions in the mailer. It is recommended that specimens arrive within 24 hours of collection.
Samples arriving on the weekend and observed holidays may be canceled.
Necessary Information
Date and time of collection and ordering healthcare professional name and phone number are required.
Specimen Required
Two separate whole blood specimens, EDTA and (possibly 2 tubes) sodium heparin, are required.
For serial monitoring, it is recommended that specimens are collected at the same time of day.
Supplies: Ambient Shipping Box-Critical Specimens Only (T668)
Specimen Type: EDTA whole blood
Container/Tube: Lavender top (EDTA)
Specimen Volume: 3 mL
Collection Instructions:
1. Send whole blood specimen in original tube. Do not aliquot.
2. Label specimen as TBBS / Quantitative Lymphocyte Subsets: T, B, and Natural Killer (NK) Cells, Blood
Additional Information: For serial monitoring, it is recommended that specimen collection be performed at the same time of day.
Specimen Type: Sodium heparin whole blood
Container/Tube: Green top (sodium heparin)
Specimen Volume: 15 mL
Collection Instructions:
1. Send whole blood specimen in original tube. Do not aliquot.
2. Label specimen as GLIC / CD8 T-Cell Immune Competence, Global, Blood
Specimen Minimum Volume
Sodium heparin whole blood: 10 mL
EDTA whole blood: 1 mL
Specimen Stability Information
Specimen Type | Temperature | Time | Special Container |
---|---|---|---|
WB Sodium Heparin | Ambient | 48 hours | GREEN TOP/HEP |
Whole Blood EDTA | Ambient | 52 hours | PURPLE OR PINK TOP/EDTA |
Reject Due To
Gross hemolysis | Reject |
Gross lipemia | Reject |
Reference Values
The appropriate age-related reference values will be provided on the report.
Day(s) Performed
Monday through Friday
CPT Code Information
86355-B cells, total count
86357-Natural killer (NK) cells, total count
86359-T cells, total count
86360-Absolute CD4/CD8 count with ratio
86356 x 2
LOINC Code Information
Test ID | Test Order Name | Order LOINC Value |
---|---|---|
GLICP | CD8 Immune Competence Panel | 80721-4 |
Result ID | Test Result Name | Result LOINC Value |
---|---|---|
30643 | IFN-g | 95204-4 |
3321 | CD45 Total Lymph Count | 27071-0 |
30644 | CD107a/b | 95203-6 |
3316 | % CD3 (T Cells) | 8124-0 |
30645 | Interpretation | 69052-9 |
3322 | CD3 (T Cells) | 8122-4 |
3319 | % CD4 (T Cells) | 8123-2 |
3325 | CD4 (T Cells) | 24467-3 |
3320 | % CD8 (T Cells) | 8101-8 |
3326 | CD8 (T Cells) | 14135-8 |
3318 | % CD19 (B Cells) | 8117-4 |
3324 | CD19 (B Cells) | 8116-6 |
4054 | % CD16+CD56 (NK cells) | 8112-5 |
4055 | CD16+CD56 (NK cells) | 20402-4 |
3327 | 4/8 Ratio | 54218-3 |
6657 | Comment | 80722-2 |
Clinical Information
CD8 T cells play an important role in the immune response to viral or intracellular infectious agents, as well as antitumor immunity and immune surveillance.
Upon activation, CD8 T cells mediate a variety of effector functions, including cytokine secretion and cytotoxicity. Interferon-gamma (IFN-gamma) is one of the early cytokines produced by CD8 T cells; it is released within a few hours of activation.(1) The cytotoxic function is mediated by the contents of the cytolytic granules.(1) Cell-surface mobilization of the cytolytic granule components, CD107a and CD107b, also known as lysosome-associated membrane proteins LAMP-1 and LAMP-2, occurs when CD8 T cells mediate their cytolytic function and degranulate.(2)
CD8 T-cell activation occurs either through the T-cell receptor peptide major histocompatibility complex or by use of a mitogen (eg, phorbol myristate acetate and the calcium ionophore ionomycin). Mitogen-mediated activation is antigen nonspecific.
Impairment of global CD8 T-cell activation (due to inherent cellular immunodeficiency or as a consequence of over-immunosuppression by therapeutic agents) results in reduced production of IFN-gamma and other cytokines, reduced cytotoxic function, and increased risk for developing infectious complications. Agents associated with over-immunosuppression include the calcineurin inhibitors (eg, cyclosporine A, FK506 [Prograf/tacrolimus], and rapamycin [sirolimus]), antimetabolites (eg, mycophenolate mofetil), and thymoglobulin.
Immunosuppression is most commonly used for allograft maintenance in solid-organ transplant recipients, to prevent graft-versus-host disease in allogeneic hematopoietic stem cell transplant patients, and to treat patients with autoimmune diseases. In these settings, reducing the risk for developing infectious complications as a result of over-immunosuppression is a clinical challenge.
Therapeutic drug monitoring is routinely used in the transplant practice to avoid overtreatment and to determine patient compliance. However, the levels of drugs measured in blood specimens do not directly correlate with the administered dose due to individual pharmacokinetic differences.(3) Furthermore, drug levels may not necessarily correlate with biological activity of the drug. Consequently, it may be beneficial to consider modification of the immunosuppression regimen based on the patient's level of functional immune competence.
This assay provides a means to evaluate over-immunosuppression within the CD8 T-cell compartment (global CD8 T-cell function). Intracellular IFN-gamma expression is a marker for CD8 T-cell activation. Surface CD107a and CD107b are markers for cytotoxic function. This test may be most useful when ordered at the end of induction immunosuppression and 2 to 3 months after maintenance immunosuppression to ensure that global CD8 T-cell function is not compromised. The test may also provide value when immunosuppression is increased to halt or prevent graft rejection, to provide information on a balance between over-immunosuppression with subsequent infectious comorbidities and under-immunosuppression with resultant graft rejection.
The absolute counts of lymphocyte subsets are known to be influenced by a variety of biological factors, including hormones, the environment, and temperature. The studies on diurnal (circadian) variation in lymphocyte counts have demonstrated progressive increase in CD4 T-cell count throughout the day, while CD8 T cells and CD19+ B cells increase between 8.30 a.m. and noon with no change between noon and afternoon. Natural killer-cell counts, on the other hand, are constant throughout the day.(4) Circadian variations in circulating T-cell counts have been shown to be negatively correlated with plasma cortisol concentration.(5-7) In fact, cortisol and catecholamine concentrations control distribution and therefore, numbers of naive versus effector CD4 and CD8 T cells.(5) It is generally accepted that lower CD4 T-cell counts are seen in the morning compared to the evening(8) and during summer compared to winter.(9) These data, therefore, indicate that timing and consistency in timing of blood collection is critical when serially monitoring patients for lymphocyte subsets.
Interpretation
Interferon-gamma (IFN-gamma) and CD107a and CD107b expression below the defined reference range are consistent with a global impairment in CD8 T-cell function, most likely due to over immunosuppression.
IFN-gamma, CD107a, and CD107b levels greater than the defined reference range are unlikely to have any clinical significance.
Cautions
This assay is specific only for CD8 T cells; it does not provide information for overall T-cell competence.
Further studies are needed to determine if, within the reference range, certain levels of interferon-gamma, CD107a, and CD107b expression confer greater or lesser degrees of risk for infectious disease.
Â
Timing and consistency in timing of blood collection are critical when serially monitoring patients for lymphocyte subsets. See Clinical Information
Method Description
CD8 T-Cell Immune Competence, Global:
Peripheral blood mononuclear cells (PBMC), which contain CD8 T cells, are stimulated with a mixture of phorbol myristate acetate and ionomycin, and with stimulatory signals derived using antibodies against the costimulatory molecules CD28/CD49d. The cells are simultaneously treated with a mixture of brefeldin A and monensin, which blocks extracellular secretion of interferon-gamma (IFN-gamma), enabling intracellular retention and detection of the protein. PBMC that have not been stimulated are used as a control to determine the background levels of IFN-gamma and CD107a and CD107b. The cells are analyzed on the Becton Dickinson FACS Canto flow cytometer, and analysis involves gating (defining) of the CD8 T cells using an antihuman CD8 antibody. Specific IFN-gamma, CD107a, and CD107b signals are determined within the "gated" CD8 T-cell population. Global CD8 T-cell immune competence is measured by the amount of IFN-gamma produced (CD8 T-cell functional activity) and surface expression of CD107a and CD107b (cytotoxicity assessment) relative to the unstimulated control and is interpreted on the basis of the reference range determined from healthy adult donors.(Unpublished Mayo method)
Quantitative Lymphocyte Subsets: T, B, and Natural Killer:
The T, B, and natural killer-cell surface marker assay uses monoclonal antibodies to identify the various membrane antigens, and flow cytometry to enumerate the number of cells expressing these differentiation antigens. CD14 is used to exclude monocytes, thereby improving accuracy and enhancing the purity of the lymphocyte population. The results are reported as the percent of lymphocytes that are total T cells (CD3+), CD3+CD4+ T cells, CD3+CD8+ T cells, natural killer (CD16+56+, CD3-), and B-lymphocytes (CD19+), and the absolute number of each cell type per mL of blood. The assay is a 7-color no-wash procedure and the absolute counts are calculated from internal bead standards. In addition, the total lymphocyte count and the CD4:CD8 ratio are reported.(Hoffman RA, Kung PC, Hansen WP, Goedstien G. Simple and rapid measurement of human T lymphocytes and their subclasses in peripheral blood. Proc Natl Acad Sci USA. 1980:77(8):4914-4917; Mandy FF, Nicholson JK, McDougal JS; CDC. Guidelines for performing single-platform absolute CD4+ T-cell determinations with CD45 gating for persons infected with human immunodeficiency virus. Centers for Disease Control and Prevention. MMWR Recomm Rep. 2003;52(RR-2):1-13)