Sign in →

Test Code RFPN Rheumatoid Factor Panel, Serum


Additional Testing Requirements


Rheumatoid factor (RF) IgA and RF IgM should be tested in combination with CCP / Cyclic Citrullinated Peptide Antibodies, IgG, Serum; or following positive or equivocal results for either RHUT / Rheumatoid Factor, Serum; or RAPAN / Rheumatoid Arthritis Panel, Serum.



Specimen Required


Supplies: Sarstedt Aliquot Tube, 5 mL (T914)

Collection Container/Tube:

Preferred: Serum gel

Acceptable: Red top

Submission Container: Plastic vial

Specimen Volume: 1 mL

Collection Instructions: Centrifuge and aliquot serum into plastic vial.


Useful For

Evaluating patients at-risk of rheumatoid arthritis (RA) or features of inflammatory arthritis

 

Differentiating the antibody isotype profile in patients positive for rheumatoid factor (RF)

 

Aid in the risk stratification of RF-seropositive patients with RA

Profile Information

Test ID Reporting Name Available Separately Always Performed
RFA Rheumatoid Factor IgA, S No Yes
RFM Rheumatoid Factor IgM, S No Yes

Method Name

Chemiluminescent Immunoassay

Reporting Name

Rheumatoid Factor Panel, S

Specimen Type

Serum

Specimen Minimum Volume

0.3 mL

Specimen Stability Information

Specimen Type Temperature Time Special Container
Serum Refrigerated (preferred) 14 days
  Frozen  28 days
  Ambient  48 hours

Reject Due To

Gross hemolysis Reject
Gross lipemia Reject
Gross icterus Reject
Heat-treated specimens Reject

Clinical Information

Rheumatoid arthritis (RA) is a chronic systemic autoimmune disease characterized by interactions between the environment, specific genetic risk factors, and the human immune system. It affects about 0.6% of the United States population with a global prevalence of 0.24%.(1) Clinically, RA is typified by progressive damage of synovial joints, inflammation, production of diverse autoantibodies, and variable extra-articular manifestations.(1,2) To facilitate early diagnosis, the American College of Rheumatology/European League Against Rheumatism 2010 RA classification criteria recommend testing for rheumatoid factors (RF) and anti-citrullinated protein antibodies (ACPA).(2) RF are autoantibodies directed against the Fc portion of immunoglobulin, while ACPA are directed against peptides and proteins containing citrulline, a modified form of the amino acid arginine.(3,4) In addition to the defined interpretations for anti-cyclic citrullinate peptide (CCP) and RF antibodies, the classification criteria also endorse the combination of specific clinical features and inflammatory markers for RA diagnosis.

 

The clinical symptoms in the early phase of RA may be nonspecific with some patients demonstrating relatively low levels of antibodies to RF or anti-CCP antibodies, which may not fulfill the diagnostic criteria for disease. In addition, some patients with clinical features of RA may test negative for criteria antibodies, a phenomenon referred to as seronegative or ACPA-negative. While alternative diagnoses may be implicated in at-risk RA patients, determination of autoantibody isotypes for RF and other RA-associated autoantibodies have been reported to improve diagnostic accuracy and/or provide prognostic clues.(5-8) Thus, determination of multiple analytes of diverse antibody isotypes in patients seropositive for RA may be useful in risk stratification for joint erosive disease and other clinical manifestations such as cardiovascular or lung involvements.(5,7,8)

 

In routine clinical laboratory evaluation for RA, RF antibodies are generally detected and quantified using IgM RF or total (isotype-nonspecific) RF immunoassays and CCP IgG antibodies with a variety of solid-phase immunoassays, such the enzyme linked immunosorbent assay, chemiluminescence immunoassay (CIA), fluorescent enzyme immunoassay (FEIA), multiplexed immunoassay using manual or automated platforms.(5,6,9,10) With respect to RF antibody measurements, it has been established that separate determination of RF IgA and RF IgM antibodies is important in RA evaluation, as severe joint erosive disease is seen more in patients with significantly elevated IgA RF than in those who are IgA RF negative.(5,8,9) However, IgA RF is generally less sensitive than IgM RF for RA, and double positivity for IgM RF and IgA RF has a higher specificity for RA than either IgM RF or IgA RF.(9) Both tests should be offered in a panel, which is not intended to replace RF tests that detect IgA, IgG and IgM autoantibodies. The relevance of IgG RF in addition to IgA or IgM RF is of limited clinical value and not available for testing on the CIA or FEIA platforms due to this clinical limitation.

Reference Values

RHEUMATOID FACTOR, IgA

Negative: <20 CU

Positive: ≥20 CU

Reference values apply to all ages.

 

RHEUMATOID FACTOR, IgM

Negative: <5.0 IU/mL

Positive: ≥5.0 IU/mL

Reference values apply to all ages.

Interpretation

The presence of rheumatoid factor (RF) IgM or IgA at abnormal levels in association with anti-citrullinated peptide/protein antibodies has high specificity for a diagnosis of rheumatoid arthritis (RA). However, the presence of RF isotypes in any combination may be found in a variety of conditions, including Sjogren syndrome and hepatitis infections.

Cautions

Some patients with rheumatoid arthritis (RA) may be negative for rheumatoid factor (RF) IgM or RF IgA but positive for RF antibody by nephelometry or anti-citrullinated peptide IgG antibodies. Negative results do not rule out a diagnosis of RA. All results must be interpreted in the appropriate clinical context and in association with other recommended laboratory tests disease evaluation. The performance characteristics of this assay have not been established for matrices other than serum.

Method Description

The Inova BIO-FLASH system uses paramagnetic beads as the solid surface upon which to bind the targeted antibody. The beads used in the rheumatoid factor (RF) IgA and IgM kits are coated with rabbit polyclonal antibodies as the capture antigen. The serum sample is diluted by the instrument and combined in a cuvette with the beads and assay buffer for 10 minutes at 37° C. The beads are magnetized, and unbound materials are washed away. Detection antibody conjugated with isoluminol is added to the cuvette for another incubation at 37° C. The RF IgA kits contain anti-human IgA conjugate, while the RF IgM kits contain anti human IgM conjugate. After an additional wash, trigger reagents are added to the cuvette to initiate a light reaction proportional to the amount of bound isoluminol conjugate. The light produced by the reaction is measured by a detector as relative light unit (RLU). The RLU is applied to the working curve on the instrument. The RF IgA is reported as chemiluminescent units (CU), while the RF IgM is reported in IU/mL traceable to WHO Reference Reagent W1066.(Package inserts: QUANTA Flash RF IgA. Inova Diagnostics, Inc; V 4, 07/2019; QUANTA Flash RF IgM. Inova Diagnostics, Inc; V 4, 07/2019)

Day(s) Performed

Wednesday

Report Available

2 to 7 days

Performing Laboratory

Mayo Clinic Laboratories in Rochester

Test Classification

This test has been cleared, approved, or is exempt by the US Food and Drug Administration and is used per manufacturer's instructions. Performance characteristics were verified by Mayo Clinic in a manner consistent with CLIA requirements.

CPT Code Information

86431 x2

LOINC Code Information

Test ID Test Order Name Order LOINC Value
RFPN Rheumatoid Factor Panel, S 106053-2

 

Result ID Test Result Name Result LOINC Value
RFM Rheumatoid Factor IgM, S 9338-5
RFA Rheumatoid Factor IgA, S 33313-8